Stratum - Einleitung
Stratum - Technik

Das OptiScan Stratum ist für 220 oder 110 V Betrieb erhältlich.

 

Mechanical Specifications
Physical size (with monitor) 1600 mm (H) x 650 mm (W) x 850 mm (D)
Total Instrument weight 150 Kg
Scanner size 270 mm (L) x 35 mm (Diameter)
Scanner weight Nominally 300 grams
Operating environment 10°C - 40°C
0% - 75% RH, non condensing
Ventilation Requirements Clearance to nearest wall 0.4 m
Heat load 2500 Watts (maximum)
 
Electrical Specifications
Power requirements 16 Amps @ 117 V AC
8 Amps @ 240 V AC
 
Laser Specifications
Laser Type Argon Ion, air cooled.
Illumination Wavelength 488 nm
Laser Life Nominally 3,000 hoursNote 1
Laser Class The instrument contains a Class 3B laser
(max output 50 mW)
Accessible Optical Output Power < 5 mW (Class 3A) at the scanner tip.
Cooling (forced air) Air cooled
 
Performance Specifications
FOV 200 µm x 200 µm (nominal)
Magnification Approximately 1000 x
Depth of imaging 200 µm typical in normal human skinNote 2
Optical resolution - lateralNote 3 < 2.5 µm
Optical resolution - axialNote 3 < 4 µm
Scan Mode X-Y, manual Z depth control
Detector Spectral Bandwidth 400 nm - 850 nm
 
Digital resolution
(acquisition mode)
Scan rate
1024 x 1024 pixels 1.4 sec/frame
1024 x 512 pixels 0.7 sec/frame
1024 x 256 pixels 0.4 sec/frame
1024 x 128 pixels 0.2 sec/frame
 
General Specifications
Guarantee/warranty 12 months guarantee against faulty workmanship and materialsNote 4
Network requirements 10 baseT or 100 baseT connection via RJ45 socket
Export file formats Proprietary Optiscan high resolution file format
PC operating system Windows NT
Monitor 19 inch touch screen monitor, 1600 x 1200 pixels
 
Applicable Standards
FDA Regulatory Class Class 1 Exempt
Classification Regulation Surgical Microscope and Accessories, No 878.4700
Electrical specifications IEC 60601-1:1988 (including amendments A1 & A2) UL 2601
AS/NZS 3200.1.0:1998
Electromagnetic Compatibility IEC 60601-1-2
Laser Safety (Class 3A) CFR21, part 1040; IEC 825-1 and AS/NZS 2211.1, 1997
 

Note 1: Laser life is subject to manufacturer's specifications and is affected by the output power at which the laser is operated.
Note 2: The depth of imaging in human skin is subject to normal biological variability and will vary according to skin type and location on the body.
Note 3: The figures quoted are measured system performance parameters and do not reflect the ultimate resolution of the optical elements.
Note 4: Refer to Optiscan's standard conditions of sale for the full conditions of warranty.
Note 5: All specifications are subject to change without notice as Optiscan continues to improve the performance of the instrument.